McSwiggen & Associates  
 

Why doesn't the standard, that I just standardize on, show up in the
Quantitative Analysis, Standard Condition window?

 
 
 
  There are three main mistakes that will cause a standard not to be displayed in the Quantitative Analysis, Standard Condition window.  
1)
  The standard was run at a different accelerating voltage than is presently set in the EOS Conditions for Quantitative Analysis.
2)
 
The standard was run using a different crystal or spectrometer than is presently set for that element in the Quantitative Analysis, Element Conditions.
3)
 
The standard was run under a different Group name than is presently selected in Quantitative Analysis. The standard Group can be changed in the Quantitative Analysis, Standard Condition window using the "STD group" button in the upper left corner.
 
 
Close window button